Validation Committee |
Document Number: |
Equ51-04 |
Effective (or Post) Date: |
19 Dec 2008 |
|
Review History |
Date of last review: |
13-Feb-2020 |
Reviewed by: |
Heidi Hanes |
|
SMILE Comments: This document is provided as an example only. It must be revised to accurately reflect your lab’s specific processes and/or specific protocol requirements. Users are directed to countercheck facts when considering their use in other applications. If you have any questions contact your SMILE representative. |
SMILE Accuracy Guidelines-Chemistry
Document Number |
210 |
|
Effective Date |
19 Dec 2008 |
|
Subject Guidelines for performing accuracy testing on a chemistry analyzer |
Page |
1 of 1 |
Supercedes |
New |
|
Review History |
Date of last review: |
19 Dec 2008 |
Reviewed by: |
Erin Gover |
Revision History |
Version # [0.0] |
Revision Date [dd/mm/yy] |
Description (notes) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SMILE Chemistry Accuracy Guidelines
ACCURACY is the true value of a substance being measured. Verification of accuracy is the process of determining that the test system is producing true, valid results.
Determine your comparison or reference method
The comparison method must be previously validated.
The comparison method must be currently performing successfully on EQA
The ideal comparison method is a similar instrument/method
Comparison to an in-house method is preferred if the in-house instrument meets the above criteria.
Samples with known values, such as proficiency testing samples or commercial standards, may be used as the reference method.
Sample Criteria
A minimum of 20 samples that cover the reportable range of the method and include points near the Medical Decision Points.
Patient, quality control, and proficiency testing materials may be used.
Testing: Run each sample in duplicate on each instrument
Ideally samples should be run within 2 hours of each other.
Duplicates should be averaged.
Data should be plotted immediately to identify and correct any outliers.
Acceptability criteria—evaluate the data using one of the options below:
Plot
the data in EP Evaluator
Two Instrument Comparison Module or a similar regression analysis
program.
Plot
the reference method on the X axis and the method being validated
on the Y axis. Enter
the Allowable Total Error (TEa) concentration and percent. Refer
to SMILE
Chemistry TE Limits
table. (Appendix 2)
Enter
the results for the reference (X) and the new (Y) methods. Evaluate
the statistics by clicking on Print Preview or on the statistics
tab. The
correlation coefficient (R) must be >0.975
If |
Then |
R <0.975 |
Data does not extend over acceptable range. More data must be evaluated over larger range. |
R >0.975 |
Proceed with step B below to evaluate acceptability. |
Using this method, 95% of the data points from the comparative method must be within Total Allowable Error limit of the reference method (refer to SMILE Chemistry TE Limits table, Appendix 2). Use one of the methods below to evaluate.
Using
a regression analysis program or spreadsheet (such as the SMILE
Accuracy Pack, Appendix
1) calculate the “Error Index” to measure the
difference between the two methods as a ratio of the Total
Allowable Error.
The
Error Index can be calculated by subtracting the reference method
data point (X) from the method being validated data point (Y) and
dividing by the Total Allowable Error (TEa). The equation is:
(Y-X)/TEa.
Use
the percent TEa or the minimum detectable difference (absolute
TEa), whichever is greater. The
absolute TEa value is calculated by converting the TEa percentage
to a decimal and multiplying by the reference method data point.
The
Error Index is measured for each X-Y pair, and must fall within
-1 and 1. If more than 5% of the specimens have an Error Index of
less than -1 or greater than 1, the accuracy experiment fails.
The
EP Evaluator
program uses a statistic called the “Error Index” to
measure the difference between the two methods as a ratio of the
Total Allowable Error.
The
Error Index can be calculated by subtracting the reference method
data point (X) from the method being validated data point (Y) and
dividing by the Total Allowable Error (TEa). The equation is:
(Y-X)/TEa.
EP
Evaluator uses the percent TEa or the minimum detectable
difference (absolute TEa), whichever is greater. The
absolute TEa value is calculated by converting the TEa percentage
to a decimal and multiplying by the reference method data point.
The
Error Index is measured for each X-Y pair, and must fall within
-1 and 1. If more than 5% of the specimens have an Error Index of
less than -1 or greater than 1, the accuracy experiment fails.
(See Appendix 3 for an example EP report).
Evaluate results using the table below:
If |
Then |
95% of Error Indices are acceptable |
Accuracy is acceptable Proceed with Linearity experiments
|
<95% of Error Indices are acceptable |
Proceed with step D below.
|
Plot the data in a regression analysis program such as the SMILE Accuracy Pack (Appendix 1) or EP Evaluator Alternate (Quantitative) Method Comparison module. Deming Regression is preferred, however if not available regular regression is acceptable provided that the Correlation Coefficient (R) is >0.975.
Plot the reference method on the X axis and the method being validated on the Y axis.
Enter the Allowable Total Error (TEa) concentration and percent. Refer to SMILE Chemistry TE Limits table (Appendix 2).
Enter Medical Decision Points
A Medical Decision Point (MDP) is the concentration of the analyte at which a medical decision is triggered. You may enter up to five values. These values are plotted on the graph, and the program computes a predicted Y value and confidence interval at each X decision point.
For many analytes, the medical decision points correspond to the lower and upper limits of the normal range. It is advisable to also include medical decision points at the DAIDS Toxicity limits and/or laboratory established critical values.
Evaluate using one of the methods described below:
If
using a regression analysis program or spreadsheet calculate and
evaluate the 95% confidence limits and error indices. The SMILE
Accuracy Pack, (Appendix
1) performs these calculations automatically.
Subtract
the X Method MDP from the low
and the high 95% confidence limits for each Medical Decision
Point. Calculate
Error Indices for each end of the 95% confidence limits for each
Medical Decision Point as follows: Divide the difference (from
step i. above) by the SMILE TE Limit absolute value. Use
the percent TEa or the minimum detectable difference, whichever is
greater.
To
calculate the absolute TEa value, convert the TEa percentage to a
decimal and multiply by the reference method data point.
All
error indices must fall within -1 and 1. An Excel spreadsheet is
provided to calculate Error Indices (See Appendix 5 SMILE Error
Index Calculator).
If
using EP Evaluator
program create the report by selecting the Print Preview button. Review
the Medical Decision Point Analysis data on page 1 of the report.
(See Appendix 4 for an example EP report) Subtract
the X Method MDP from the low
and the high 95% confidence limits for each Medical Decision
Point.
Calculate
Error Indices for each end of the 95% confidence limits for each
Medical Decision Point as follows: Divide the difference (from
step c above) by the SMILE TE Limit absolute value. Use
the percent TEa or the minimum detectable difference, whichever
is greater.
To
calculate the absolute TEa value, convert the TEa percentage to
a decimal and multiply by the reference method data point.
All
error indices must fall within -1 and 1. An Excel spreadsheet is
provided to calculate Error Indices (See Appendix 5 SMILE Error
Index Calculator).
Evaluate acceptability using the table below:
If |
Then |
All error indices fall between -1 and 1 |
Proceed with step 6 below. |
Any error indices fall outside -1 and 1 |
Consider possible causes of inaccuracy. Troubleshoot before repeating accuracy studies. Consult SMILE for assistance. |
Six Sigma method acceptability evaluation
Plot the data in a Six Sigma analysis program such as the SMILE Accuracy Pack (Appendix 1) or EP Evaluator Six Sigma Metrics module.
Enter the Allowable Total Error (TEa) concentration and percent. Refer to SMILE Chemistry TE Limits table.
Enter Medical Decision Points.
Enter the slope and intercept calculated in step D above.
If using EP Evaluator, use the Deming Regression slope and intercept calculated by the program.
If EP Evaluator is not used, it is acceptable to use slope and intercept calculated using regular regression.
Enter the Means and %CVs from your Between Day Precision experiments.
If using EP Evaluator create the report by selecting the Print Preview button. (See Appendix 6 for an example EP report).
The SMILE Accuracy Pack and EP Evaluator calculate a Sigma metric for each Medical Decision point.
Sigma metric=(TEa-Systematic Error)/Random Error
Evaluate acceptability at each MDP using the table below:
If |
Then |
The Sigma metric less than 2.0 |
The method has unacceptable performance and does not meet your requirement for quality, even when the method is working properly. It is not acceptable for routine operation. |
The Sigma metric is between 2.0-3.0 |
The method has marginal performance and provides the necessary quality when everything is working correctly. This method will require: 4-8 controls per run well-trained operators reduced rotation of personnel more aggressive preventive maintenance careful monitoring of patient test results continual efforts to improve method performance |
If the Sigma metric is between 3.0-4.0 |
The method has fair performance and meets your requirement for quality and can be well-managed in routine operation. This method will require 2 control measurements per run using standard Westgard QC rules. |
If the Sigma metric is between 4.0-6.0 |
The method has good performance and is clearly acceptable and can be well-managed in routine operation with only 2 control measurements per run using standard Westgard QC rules. |
If the Sigma metric is >6.0 |
The method has Six Sigma performance and is even easier to manage and control. |
References
GCLP Guidelines, Standard for Performance Specifications.
Clinical and Laboratory Standards Institute (CLSI). User Verification of Performance for Precision and Trueness: Approved Guideline-Second Edition. CLSI document EP15-A2 (ISBN 1-56238-574-7). Clinical and Laboratory Standards Institutes, 940 West Valley Road, Suite 100, Wayne, Pennsylvania 19098-1898 USA.
NCCLS. (Currently CLSI) Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline—Second Edition. NCCLS document EP5-A2 (ISBN 1-56238-542-9). NCCLS, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA.
Clinical and Laboratory Standards Institute (CLSI). Method Comparison and Bias Estimation Using Patient Samples: approved Guidelines- Second Edition. CLSI document EP9-A2 (ISBN 1-56238-472-4). Clinical and Laboratory Standards Institutes, 940 West Valley Road, Suite 100, Wayne, Pennsylvania 19098-1898 USA.
Clinical and Laboratory Standards Institute (CLSI).Preliminary Evaluation of Quantitative Clinical Laboratory Measurement Procedure: Approved Guidline – Third Edition. CLSI document EP10-A3 (ISBN 1-56238-622-0). ). Clinical and Laboratory Standards Institutes, 940 West Valley Road, Suite 100, Wayne, Pennsylvania 19098-1898 USA.
Clinical and Laboratory Standards Institute (CLSI).Estimation of Total analytical Error for Clinical Laboratory Methods. Approved 2003. CLSI document EP21-A (ISBN 1-56238-502-xX. Clinical and Laboratory Standards Institutes, 940 West Valley Road, Suite 100, Wayne, Pennsylvania 19098-1898 USA.
EP Evaluator Release 8, David G. Rhoads Associates Inc., www.dgrhoads.com.
James O. Westgard, Online Validation Training, Westgard QC, Inc. www.westgard.com, Sections 11-Determining Bias,12- Estimating Trueness, and 13- Judging Method Acceptability.
2 CERTIFICATE OF CORPORATE AUTHORITY [NAME
2 MEETING OF NATIONAL AUTHORITIES OEASER KXXXIX
6 ARTICLE 6B INTERIM CASINO AUTHORIZATION 5129512
Tags: number, equ5104, document, validation, committee, post), author, effective